At Poretech Analytical Testing Services, we are committed to helping our customers obtain accurate pore structure information of their materials. Since no single measurement technique works for every sample, it is critical to identify the suitable measurement technique.
Backed by a team of in-house material scientists, we begin by discussing your application, identifying the suitable techniques and providing the most comprehensive results. While Poretech provides you with a detailed report, our Application Scientists are always available to discuss and help interpret your results.
Poretech Analytical Testing Services Division can accommodate a wide variety of samples, including non-wovens, filtration membranes, ceramics, barrier materials, specialty textiles, battery separators, hydrophobic and hydrophilic materials, and many more with different shape, sizes and test conditions. The sample size and consistency requirements may vary with the type of tests and materials to be analyzed. If the test result is urgent, we can also offer priority testing service to analyze your samples and have the results sent to you within 1-2 business days as well.
Testing Method | Item No. | Testing Results |
---|---|---|
Gas-Liquid | GL-01 | Bubble Point |
Gas-Liquid | GL-02 | Bubble Point, Mean Flow Pore Size and Pore Distribution |
Gas-Liquid | -GP | Gas Permeability |
Gas-Liquid | -H | Hollow Fiber Test (or other non-flat sheet sample) |
Liquid-Liquid | LL-01 | Bubble Point |
Liquid-Liquid | LL-02 | Bubble Point, Mean Flow Pore Size and Pore Distribution |
Liquid-Liquid | -LP | Liquid Permeability |
Liquid-Liquid | -H | Hollow Fiber Test (or other non-flat sheet sample) |
Water Intrusion (Hydrophobic Material) | WI-01 | Pore Volume, Pore Size Distribution |
Water Intrusion (Hydrophobic Material) | WI-02 | Porosity % |
Testing Item | Measuring methods | Analysis methods | Measuring range |
---|---|---|---|
Contact Angle | Sessile drop (static and dynamic) | θ/2, tangent, circle fitting, ellipse fitting | 0.1 to 180° |
Surface Tension | Wilhelmy plate and du Noüy ring method | Surface tension, Interfacial tension, Lamella length, Sedimentation behavior | 1 to 1000mN/m |
Testing Result | Testing Standard | Required Sample Q'ty | Required Performance of Type I mask | Required Performance of Type II mask | Required Performance of Type IIR mask |
---|---|---|---|---|---|
Bacterial Filtration Efficiency (BFE) | EN 14683 Annex B | 10 pcs | ≥ 95 % | ≥ 98 % | ≥ 98 % |
Differential Pressure (DP) | EN 14683 Annex C | 10 pcs | < 40 Pa/cm² | < 40 Pa/cm² | < 60 Pa/cm² |
Splash Resistance Pressure | ISO 22609 | 18 pcs | Not required | Not required | > 16.0 kPa |
Microbial Cleanliness | EN 14683 / EN ISO 11737-1 | 10 pcs | < 30 cfu/g | < 30 cfu/g | < 30 cfu/g |
Initial Efficiency and Resistance | 42 CFR Part 84.182 | 15 pcs | ------ | ------ | ------ |
Testing Result | Testing Standard | Required Sample Q'ty | Required Performance of Level 1 | Required Performance of Level 2 | Required Performance of Level 3 |
---|---|---|---|---|---|
Bacterial Filtration Efficiency (BFE) | ASTM F2100 9.1 / ASTM F2101 | 10 pcs | ≥ 95 % | ≥ 98 % | ≥ 98 % |
Virus Filtration Efficiency (VFE) | ASTM F2101 modified | 7 pcs | ------ | ------ | ------ |
Differential Pressure (DP) | ASTM F2100 9.2 / EN 14683 Annex C | 10 pcs | < 5.0 mm H2O/cm² | < 6.0 mm H2O/cm² | < 6.0 mm H2O/cm² |
Particulate Filtration Efficiency (PFE) | ASTM F200 9.3 / ASTM F2299 | 10 pcs | ≥ 95 % | ≥ 98 % | ≥ 98 % |
Resistance to Penetration by Synthetic Blood (min. pressure in mmHg for pass result) | ASTM F2100 9.4 / ASTM F1862 | 40 pcs | Pass at 80 mmHg | Pass at 120 mmHg | Pass at 160 mmHg |
Flammability | ASTM F2100 9.5 / 16 CFR PART 1610 | 15 pcs | Class 1 (≥ 3.5 seconds) | Class 1 (≥ 3.5 seconds) | Class 1 (≥ 3.5 seconds) |